5d687f7154
전체 사용 흐름 (연결→정렬→측정→배뇨기록) 단계별 안내 Placement 3-step 절차, Auto/Spot 측정, 설정, 트러블슈팅 포함 Co-Authored-By: Claude Opus 4.6 (1M context) <noreply@anthropic.com>
9.0 KiB
9.0 KiB
VesiScan-Basic User Guide
App Version: VBTAND0101
1. Getting Started
First Launch
- Install the app and open it
- Complete the onboarding screens
- Register your information (name, age, height, weight)
- Set a 4-digit PIN
- You will arrive at the Home screen
Home Screen
- Tap Start to begin
- The app version is displayed below "Smart Bladder Monitoring"
- To enable Developer Mode: tap the Bladdy character 3 times quickly (within 1.5 seconds)
2. Connecting the Device
- Tap Start on the Home screen
- The app will scan for nearby VesiScan devices
- When your device appears in the list, tap it to connect
- Complete the Bluetooth pairing when the system dialog appears
- Do NOT press Cancel — the app will not proceed without pairing
- Once paired, you will move to the Personalization screen
Personalization
- Set your posture (standing/sitting/lying)
- Set the sensor position
- Set the Max Volume (default: 500 mL)
- Tap Next to proceed to Sensor Alignment
3. Sensor Alignment (Placement)
This step ensures the sensor is positioned correctly over your bladder for accurate measurement.
Step 1: Vertical Alignment
- Place the VesiScan-Basic device just above your pubic bone
- Apply ultrasound gel between the sensor and your skin
- Tap the Start Alignment button
- The app will begin scanning and provide direction:
- "Raise the sensor up" — slide the device upward toward your navel
- "Lower the sensor down" — slide the device downward toward your pubic bone
- "Raise the sensor slightly" / "Lower the sensor slightly" — make small adjustments
- Move the sensor slowly, 1-2mm at a time
- When the vertical position is correct, the app will show Step 2/3
The hint text will show animated dots (e.g., "Raise the sensor up.", "Raise the sensor up..", "Raise the sensor up...") to indicate the system is actively scanning.
Step 2: Lateral Alignment
- The app will first show "Checking left/right balance..." for a few seconds
- Then it will guide you:
- "Move left" — slide the device to the left
- "Move right" — slide the device to the right
- Adjust until both lateral sensors (CH4 and CH5) detect the bladder equally
- When balanced, the app will show Step 3/3
Step 3: Final Check (Green Zone)
- The app checks overall alignment quality
- If everything is optimal, the hint will show "Optimal position!"
- The screen will hold this state for 7 seconds to confirm stability
- Tap Alignment Complete to proceed to the measurement screen
If Alignment is Lost
- If the sensor shifts significantly after reaching Optimal:
- The app will show "Position lost. Place sensor above pubic bone, then press Start"
- The scanning will stop automatically
- Re-place the sensor and tap Start Alignment again
If the Sensor is Detached
- If the sensor loses contact with skin:
- The app will show "Sensor detached. Place above pubic bone, then press Start"
- Re-attach the sensor with gel and tap Start Alignment
Tips
- Use plenty of ultrasound gel — dry contact causes poor readings
- Keep the sensor flat against the skin — tilting reduces accuracy
- Move slowly — fast movements cause unstable readings
- You can tap Skip to go directly to measurement without alignment
4. Measurement Screen
After alignment, you will see the main measurement screen with a donut chart.
Screen Layout
[Placement] [Battery] [Catheters] [Settings]
Enjoy your day!
┌─────────────────┐
│ Donut Chart │
│ │
│ [Bladdy] │
│ 325 mL │
└─────────────────┘
Current Measurement | 65%
280 mL | 325/500 mL
[Void/Catheterize] [Auto] [Spot]
Understanding the Display
| Element | Description |
|---|---|
| Donut Chart | Fills based on max measured volume / max volume setting |
| Center Value | Maximum measured volume this session (updates every 5 seconds) |
| Current Measurement | Latest trimmed mean value (Auto) or last Spot result |
| Fill Card | Top: fill percentage, Bottom: max measured / bladder max volume |
| Battery Icon | Device battery level with visual indicator |
| Catheter Count | Remaining catheters (tap to add more) |
5. Auto Measurement
- Tap the Auto button (orange)
- The device will measure continuously every ~1.5 seconds
- For the first 5 measurements, the display will show "—" (collecting data)
- After 5+ measurements, the trimmed mean value will be displayed
- The donut chart will update every 5 seconds
- Tap Stop (red) to end Auto measurement
Auto Measurement Failure
- If measurements fail 5 times in a row (sensor shifted or lost contact):
- A dialog will appear: "Sensor Position Issue"
- Tap Go to Alignment to re-align the sensor
- Or tap Dismiss to stay on the measurement screen
6. Spot Measurement
- Tap the Spot button
- The device will take 5 consecutive measurements (~4 seconds total)
- A spinner will show during measurement
- The result (trimmed mean of 5 readings) will be displayed immediately
- The donut chart and Current Measurement card will update
During Spot measurement, the Auto and Void buttons are disabled.
7. Void / Catheterize
- Tap the Void / Catheterize button
- The current measurement is recorded to the voiding diary
- The catheter count decreases by 1
- The bladder level resets to 0
- A toast message "Voiding recorded" will appear
Managing Catheters
- The catheter count is shown at the top of the screen (hospital icon + number)
- Tap the catheter count to add more catheters
- Enter the number and tap Add
- Default starting count: 15 catheters
8. Settings
Tap the gear icon at the top right to open settings.
General Settings (always visible)
| Setting | Description | Default |
|---|---|---|
| Max Volume | Maximum bladder volume for fill calculation | 500 mL |
| Catheter Threshold | Alert level for catheterization | Level 7 |
Developer Settings (Developer Mode only)
| Setting | Description |
|---|---|
| Threshold | Otsu (auto) or manual echo threshold |
| DPS | Distance per sample (mm) |
| Detection | Method A / B / C |
| BV | Frustum / V41 (volume calculation method) |
| SG Filter | Savitzky-Golay noise filter on/off |
| Post Max | Maximum sample index for wall detection (filters floor reflections) |
Saving Settings
- Adjust values using sliders and toggles
- Tap Save to apply and close
- Tap Cancel to close without changes (note: slider changes apply in real-time)
9. Navigation
| Action | Result |
|---|---|
| Placement button (top left) | Go to Sensor Alignment |
| Settings button (top right) | Open/close settings panel |
| Back button (Android) | Go to previous screen |
| Back button on Home | "Exit App?" confirmation dialog |
10. Troubleshooting
| Problem | Solution |
|---|---|
| "Place sensor above pubic bone" won't go away | Make sure the sensor is attached with gel, then press Start Alignment |
| Alignment keeps showing "Raise/Lower" | Move the sensor very slowly, 1-2mm at a time |
| Auto measurement shows "—" | Wait for at least 5 measurements to accumulate |
| Donut chart value doesn't change | Max volume updates every 5 seconds |
| Spot measurement takes too long | Each Spot takes ~4 seconds (5 readings) |
| "Sensor Position Issue" dialog appears | The sensor may have shifted — go to Alignment |
| Bluetooth connection lost | The app will attempt auto-reconnect (up to 5 times) |
| App freezes on measurement screen | Check BLE connection, try disconnecting and reconnecting |
| Pairing dialog appears but app won't proceed | You must tap Pair — canceling will block the connection |
11. LED Indicators (Device)
| LED State | Meaning |
|---|---|
| Green blink (0.5s) | Bluetooth scanning |
| Green on 1s / off 3s | Device idle (not attached) |
| Orange blink (1s) | Sensor Alignment in progress |
| Green solid | Alignment complete (Optimal position) |
| Orange fast blink (3Hz) | Error detected |
| Orange solid | Charging |
| Blue solid | Charging complete |
12. Important Notes
- Always use ultrasound gel between the sensor and skin
- Keep the sensor still during measurement — movement reduces accuracy
- Do not use the device while charging
- The app keeps the screen on during use to prevent BLE disconnection
- Measurement logs are automatically saved to your device:
- BLE logs:
Downloads/VesiScan_BLE_*.log - ADC data:
Downloads/VesiScan_ADC/*.csv
- BLE logs:
Contact
For technical support, contact:
- App Development: dwjang
- Algorithm: Charles KWON / eunji.won
- Medithings Co., Ltd. — https://medithings.net