- Pipeline Summary: GREEN 상수 (CV preset별, LR 0.20, 7s hold, 3fail exit), 힌트 텍스트 표 추가, Spot 5회/Auto 10윈도우, 태블릿 fontScale, 향후 과제 갱신 (IMU 포팅 예정) - USER_GUIDE: 전체 재작성 — Start flow, 힌트 화살표(↑↓←→), 페어링 다이얼로그, Void/Catheterization, Single Scan/Auto Scan, Checking/Final check/In position! 텍스트 반영 - ALGORITHM_COMPARISON: Placement 섹션 현행화 — Start 버튼 필수, phase protection, 화살표 항상 갱신, GREEN 7s+3fail Co-Authored-By: Claude Opus 4.6 (1M context) <noreply@anthropic.com>
256 lines
8.9 KiB
Markdown
256 lines
8.9 KiB
Markdown
# VesiScan-Basic User Guide
|
|
|
|
## App Version: VBTAND0101
|
|
## Last Updated: 2026-05-11
|
|
|
|
---
|
|
|
|
## 1. Getting Started
|
|
|
|
### First Launch
|
|
1. Install the app and open it
|
|
2. Complete the onboarding screens
|
|
3. Register your information (name, age, height, weight)
|
|
4. Set a 4-digit PIN
|
|
5. You will arrive at the **Home** screen
|
|
|
|
### Home Screen
|
|
- Tap **Start** to begin
|
|
- The app version is displayed below "Smart Bladder Monitoring"
|
|
- To enable Developer Mode: tap the Bladdy character **3 times quickly** (within 1.5 seconds)
|
|
|
|
---
|
|
|
|
## 2. Connecting the Device
|
|
|
|
1. Tap **Start** on the Home screen
|
|
2. The app will scan for nearby VesiScan devices (sorted alphabetically)
|
|
3. When your device appears in the list, tap it to connect
|
|
4. **Complete the Bluetooth pairing** when the system dialog appears
|
|
- A "Pairing Required" dialog will guide you through the process
|
|
- Tap **Pair** on the system Bluetooth dialog — canceling will block the connection
|
|
- The app waits up to 30 seconds for pairing to complete
|
|
5. Once paired, you will move to the Personalization screen
|
|
|
|
### Personalization
|
|
- Set your **Max Volume** (200-800 mL, default: 500 mL)
|
|
- Tap **Continue** to proceed to Sensor Alignment
|
|
|
|
---
|
|
|
|
## 3. Sensor Alignment (Placement)
|
|
|
|
This step ensures the sensor is positioned correctly over your bladder for accurate measurement.
|
|
|
|
### Before Starting
|
|
- Place VesiScan above the pubic bone
|
|
- Apply ultrasound gel between the sensor and your skin
|
|
- Tap the **Start Alignment** button (large orange button)
|
|
|
|
### Step 1/3: Vertical Alignment
|
|
|
|
The app will begin scanning and provide direction:
|
|
- **"Slide up ↑"** — slide the device upward toward your navel
|
|
- **"Slide down ↓"** — slide the device downward toward your pubic bone
|
|
- **"Slide up slightly ↑"** / **"Slide down slightly ↓"** — make small adjustments
|
|
- Move the sensor **slowly, 1-2mm at a time**
|
|
- When the vertical position is correct, the app will show **Step 2/3**
|
|
|
|
> The hint text will show animated dots (e.g., "Slide up ↑.", "Slide up ↑..", "Slide up ↑...") to indicate the system is actively scanning.
|
|
|
|
### Step 2/3: Lateral Alignment
|
|
|
|
1. The app will first show **"Checking..."** for a few seconds
|
|
2. Then it will guide you:
|
|
- **"Slide left ←"** — slide the device to the left
|
|
- **"Slide right →"** — slide the device to the right
|
|
3. Adjust until both lateral sensors (CH4 and CH5) detect the bladder equally
|
|
4. When balanced, the app will show **Step 3/3**
|
|
|
|
### Step 3/3: Final Check (Green Zone)
|
|
|
|
1. The app checks overall alignment quality (CV + LR deviation)
|
|
2. If everything is optimal, the hint will show **"In position!"**
|
|
3. The screen background turns green and holds this state for **7 seconds** to confirm stability
|
|
4. The **Start Scanning** button (green) becomes active
|
|
5. Tap **Start Scanning** to proceed to the measurement screen
|
|
|
|
### If Alignment is Lost
|
|
- If the sensor shifts significantly after reaching "In position!":
|
|
- The app allows up to **3 consecutive failures** before exiting GREEN
|
|
- After 3 failures: **"Position lost. Restart above pubic bone."**
|
|
- The scanning will stop automatically
|
|
- Re-place the sensor and tap **Start Alignment** again
|
|
|
|
### If the Sensor is Detached
|
|
- If the sensor loses contact with skin:
|
|
- The app will show **"Sensor detached. Restart above pubic bone."**
|
|
- Re-attach the sensor with gel and tap **Start Alignment**
|
|
|
|
### Tips
|
|
- Use **plenty of ultrasound gel** — dry contact causes poor readings
|
|
- Keep the sensor **flat against the skin** — tilting reduces accuracy
|
|
- Move **slowly** — fast movements cause unstable readings
|
|
- You can tap **Skip** to go directly to measurement without alignment
|
|
|
|
---
|
|
|
|
## 4. Measurement Screen
|
|
|
|
After alignment, you will see the main measurement screen with a donut chart.
|
|
|
|
### Screen Layout
|
|
|
|
```
|
|
[Placement] [Battery] [Catheters] [Settings]
|
|
|
|
Enjoy your day!
|
|
|
|
┌─────────────────┐
|
|
│ Donut Chart │
|
|
│ │
|
|
│ [Bladdy] │
|
|
│ 325 mL │
|
|
└─────────────────┘
|
|
|
|
Current Measurement | 65%
|
|
280 mL | 325/500 mL
|
|
|
|
[Void/Catheterization] [Auto Scan] [Single Scan]
|
|
```
|
|
|
|
### Understanding the Display
|
|
|
|
| Element | Description |
|
|
|---------|-------------|
|
|
| **Donut Chart** | Fills based on max measured volume / max volume setting |
|
|
| **Center Value** | Maximum measured volume this session (updates every 5 seconds) |
|
|
| **Current Measurement** | Latest trimmed mean value (Auto) or last Spot result |
|
|
| **Fill Card** | Top: fill percentage, Bottom: max measured / bladder max volume |
|
|
| **Battery Icon** | Device battery level with visual indicator |
|
|
| **Catheter Count** | Remaining catheters (tap to add more) |
|
|
|
|
---
|
|
|
|
## 5. Auto Scan
|
|
|
|
1. Tap the **Auto Scan** button (orange)
|
|
2. The device will measure continuously (minimum 800ms interval)
|
|
3. For the first 5 measurements, the display will show **"—"** (collecting data)
|
|
4. After 5+ measurements, the **trimmed mean** value will be displayed (10-sample window, top/bottom excluded)
|
|
5. The donut chart will update every 5 seconds
|
|
6. Tap **Stop Scan** (red) to end Auto measurement
|
|
|
|
### Auto Measurement Failure
|
|
- If 2+ channels fail **5 times in a row** (sensor shifted or lost contact):
|
|
- A dialog will appear: **"Position lost. Restart from alignment"**
|
|
- Tap **Realign** to re-align the sensor
|
|
- Or tap **Dismiss** to stay on the measurement screen
|
|
|
|
---
|
|
|
|
## 6. Single Scan (Spot)
|
|
|
|
1. Tap the **Single Scan** button
|
|
2. The device will take **5 consecutive measurements** (~4 seconds total, max 8 attempts)
|
|
3. A **spinner** will show during measurement
|
|
4. The result (trimmed mean of 5 readings) will be displayed immediately
|
|
5. The donut chart and Current Measurement card will update
|
|
|
|
> During Single Scan, the Auto Scan and Void buttons are disabled.
|
|
|
|
---
|
|
|
|
## 7. Void / Catheterization
|
|
|
|
1. Tap the **Void / Catheterization** button
|
|
2. The current measurement is recorded to the voiding diary
|
|
3. The catheter count decreases by 1
|
|
4. The bladder level resets to 0
|
|
5. A toast message **"Recorded"** will appear
|
|
|
|
### Managing Catheters
|
|
- The catheter count is shown at the top of the screen (hospital icon + number)
|
|
- **Tap the catheter count** to add more catheters
|
|
- Enter the number and tap **Add**
|
|
|
|
---
|
|
|
|
## 8. Settings
|
|
|
|
Tap the **gear icon** at the top right to open settings.
|
|
|
|
### General Settings (always visible)
|
|
| Setting | Description | Default |
|
|
|---------|-------------|---------|
|
|
| Max Volume | Maximum bladder volume for fill calculation | 500 mL |
|
|
| Catheter Threshold | Alert level for catheterization | Level 7 |
|
|
|
|
### Developer Settings (Developer Mode only)
|
|
| Setting | Description |
|
|
|---------|-------------|
|
|
| Threshold | Otsu (auto) or manual echo threshold |
|
|
| DPS | Distance per sample (mm) |
|
|
| Detection | Method A / B / C |
|
|
| BV | Frustum / V41 (volume calculation method) |
|
|
| SG Filter | Savitzky-Golay noise filter on/off |
|
|
| Post Max | Maximum sample index for wall detection (filters floor reflections) |
|
|
|
|
---
|
|
|
|
## 9. Navigation
|
|
|
|
| Action | Result |
|
|
|--------|--------|
|
|
| **Placement button** (top left) | Go to Sensor Alignment |
|
|
| **Settings button** (top right) | Open/close settings panel |
|
|
| **Back button** (Android) | Go to previous screen |
|
|
| **Back button on Home** | "Exit App?" confirmation dialog |
|
|
|
|
---
|
|
|
|
## 10. Troubleshooting
|
|
|
|
| Problem | Solution |
|
|
|---------|----------|
|
|
| "Place VesiScan above the pubic bone" won't go away | Make sure the sensor is attached with gel, then press **Start Alignment** |
|
|
| Alignment keeps showing "Slide up/down" | Move the sensor **very slowly**, 1-2mm at a time |
|
|
| Auto measurement shows "—" | Wait for at least 5 measurements to accumulate |
|
|
| Donut chart value doesn't change | Max volume updates every 5 seconds |
|
|
| Single Scan takes too long | Each scan takes ~4 seconds (5 readings) |
|
|
| "Position lost" dialog appears | The sensor may have shifted — go to Alignment |
|
|
| Bluetooth connection lost | The app will attempt auto-reconnect (up to 5 times) |
|
|
| Pairing dialog appears but app won't proceed | You must tap **Pair** — canceling will block the connection |
|
|
|
|
---
|
|
|
|
## 11. LED Indicators (Device)
|
|
|
|
| LED State | Meaning |
|
|
|-----------|---------|
|
|
| State 0 | OFF (default) |
|
|
| State 4 | Sensor detached warning |
|
|
| State 5 | Alignment searching (orange) |
|
|
| State 6 | Alignment complete / In position (green) |
|
|
|
|
---
|
|
|
|
## 12. Important Notes
|
|
|
|
- **Always use ultrasound gel** between the sensor and skin
|
|
- **Keep the sensor still** during measurement — movement reduces accuracy
|
|
- **Do not use the device while charging**
|
|
- The app keeps the screen on during use to prevent BLE disconnection
|
|
- Measurement logs are automatically saved to your device:
|
|
- BLE logs: `Downloads/VesiScan_BLE_*.log`
|
|
- ADC data: `Downloads/VesiScan_ADC/*.csv`
|
|
|
|
---
|
|
|
|
## Contact
|
|
|
|
For technical support, contact:
|
|
- App Development: dwjang
|
|
- Algorithm: Charles KWON / eunji.won
|
|
- Medithings Co., Ltd. — https://medithings.net
|