# VesiScan-Basic User Guide
## App Version: VBTAND0101
---
## 1. Getting Started
### First Launch
1. Install the app and open it
2. Complete the onboarding screens
3. Register your information (name, age, height, weight)
4. Set a 4-digit PIN
5. You will arrive at the **Home** screen
### Home Screen
- Tap **Start** to begin
- The app version is displayed below "Smart Bladder Monitoring"
- To enable Developer Mode: tap the Bladdy character **3 times quickly** (within 1.5 seconds)
---
## 2. Connecting the Device
1. Tap **Start** on the Home screen
2. The app will scan for nearby VesiScan devices
3. When your device appears in the list, tap it to connect
4. **Complete the Bluetooth pairing** when the system dialog appears
- Do NOT press Cancel — the app will not proceed without pairing
5. Once paired, you will move to the Personalization screen
### Personalization
- Set your **posture** (standing/sitting/lying)
- Set the **sensor position**
- Set the **Max Volume** (default: 500 mL)
- Tap **Next** to proceed to Sensor Alignment
---
## 3. Sensor Alignment (Placement)
This step ensures the sensor is positioned correctly over your bladder for accurate measurement.
### Step 1: Vertical Alignment
1. **Place the VesiScan-Basic device just above your pubic bone**
2. Apply ultrasound gel between the sensor and your skin
3. Tap the **Start Alignment** button
4. The app will begin scanning and provide direction:
- **"Raise the sensor up"** — slide the device upward toward your navel
- **"Lower the sensor down"** — slide the device downward toward your pubic bone
- **"Raise the sensor slightly"** / **"Lower the sensor slightly"** — make small adjustments
5. Move the sensor **slowly, 1-2mm at a time**
6. When the vertical position is correct, the app will show **Step 2/3**
> The hint text will show animated dots (e.g., "Raise the sensor up.", "Raise the sensor up..", "Raise the sensor up...") to indicate the system is actively scanning.
### Step 2: Lateral Alignment
1. The app will first show **"Checking left/right balance..."** for a few seconds
2. Then it will guide you:
- **"Move left"** — slide the device to the left
- **"Move right"** — slide the device to the right
3. Adjust until both lateral sensors (CH4 and CH5) detect the bladder equally
4. When balanced, the app will show **Step 3/3**
### Step 3: Final Check (Green Zone)
1. The app checks overall alignment quality
2. If everything is optimal, the hint will show **"Optimal position!"**
3. The screen will hold this state for **7 seconds** to confirm stability
4. Tap **Alignment Complete** to proceed to the measurement screen
### If Alignment is Lost
- If the sensor shifts significantly after reaching Optimal:
- The app will show **"Position lost. Place sensor above pubic bone, then press Start"**
- The scanning will stop automatically
- Re-place the sensor and tap **Start Alignment** again
### If the Sensor is Detached
- If the sensor loses contact with skin:
- The app will show **"Sensor detached. Place above pubic bone, then press Start"**
- Re-attach the sensor with gel and tap **Start Alignment**
### Tips
- Use **plenty of ultrasound gel** — dry contact causes poor readings
- Keep the sensor **flat against the skin** — tilting reduces accuracy
- Move **slowly** — fast movements cause unstable readings
- You can tap **Skip** to go directly to measurement without alignment
---
## 4. Measurement Screen
After alignment, you will see the main measurement screen with a donut chart.
### Screen Layout
```
[Placement] [Battery] [Catheters] [Settings]
Enjoy your day!
┌─────────────────┐
│ Donut Chart │
│ │
│ [Bladdy] │
│ 325 mL │
└─────────────────┘
Current Measurement | 65%
280 mL | 325/500 mL
[Void/Catheterize] [Auto] [Spot]
```
### Understanding the Display
| Element | Description |
|---------|-------------|
| **Donut Chart** | Fills based on max measured volume / max volume setting |
| **Center Value** | Maximum measured volume this session (updates every 5 seconds) |
| **Current Measurement** | Latest trimmed mean value (Auto) or last Spot result |
| **Fill Card** | Top: fill percentage, Bottom: max measured / bladder max volume |
| **Battery Icon** | Device battery level with visual indicator |
| **Catheter Count** | Remaining catheters (tap to add more) |
---
## 5. Auto Measurement
1. Tap the **Auto** button (orange)
2. The device will measure continuously every ~1.5 seconds
3. For the first 5 measurements, the display will show **"—"** (collecting data)
4. After 5+ measurements, the **trimmed mean** value will be displayed
5. The donut chart will update every 5 seconds
6. Tap **Stop** (red) to end Auto measurement
### Auto Measurement Failure
- If measurements fail **5 times in a row** (sensor shifted or lost contact):
- A dialog will appear: **"Sensor Position Issue"**
- Tap **Go to Alignment** to re-align the sensor
- Or tap **Dismiss** to stay on the measurement screen
---
## 6. Spot Measurement
1. Tap the **Spot** button
2. The device will take **5 consecutive measurements** (~4 seconds total)
3. A **spinner** will show during measurement
4. The result (trimmed mean of 5 readings) will be displayed immediately
5. The donut chart and Current Measurement card will update
> During Spot measurement, the Auto and Void buttons are disabled.
---
## 7. Void / Catheterize
1. Tap the **Void / Catheterize** button
2. The current measurement is recorded to the voiding diary
3. The catheter count decreases by 1
4. The bladder level resets to 0
5. A toast message **"Voiding recorded"** will appear
### Managing Catheters
- The catheter count is shown at the top of the screen (hospital icon + number)
- **Tap the catheter count** to add more catheters
- Enter the number and tap **Add**
- Default starting count: **15 catheters**
---
## 8. Settings
Tap the **gear icon** at the top right to open settings.
### General Settings (always visible)
| Setting | Description | Default |
|---------|-------------|---------|
| Max Volume | Maximum bladder volume for fill calculation | 500 mL |
| Catheter Threshold | Alert level for catheterization | Level 7 |
### Developer Settings (Developer Mode only)
| Setting | Description |
|---------|-------------|
| Threshold | Otsu (auto) or manual echo threshold |
| DPS | Distance per sample (mm) |
| Detection | Method A / B / C |
| BV | Frustum / V41 (volume calculation method) |
| SG Filter | Savitzky-Golay noise filter on/off |
| Post Max | Maximum sample index for wall detection (filters floor reflections) |
### Saving Settings
- Adjust values using sliders and toggles
- Tap **Save** to apply and close
- Tap **Cancel** to close without changes (note: slider changes apply in real-time)
---
## 9. Navigation
| Action | Result |
|--------|--------|
| **Placement button** (top left) | Go to Sensor Alignment |
| **Settings button** (top right) | Open/close settings panel |
| **Back button** (Android) | Go to previous screen |
| **Back button on Home** | "Exit App?" confirmation dialog |
---
## 10. Troubleshooting
| Problem | Solution |
|---------|----------|
| "Place sensor above pubic bone" won't go away | Make sure the sensor is attached with gel, then press **Start Alignment** |
| Alignment keeps showing "Raise/Lower" | Move the sensor **very slowly**, 1-2mm at a time |
| Auto measurement shows "—" | Wait for at least 5 measurements to accumulate |
| Donut chart value doesn't change | Max volume updates every 5 seconds |
| Spot measurement takes too long | Each Spot takes ~4 seconds (5 readings) |
| "Sensor Position Issue" dialog appears | The sensor may have shifted — go to Alignment |
| Bluetooth connection lost | The app will attempt auto-reconnect (up to 5 times) |
| App freezes on measurement screen | Check BLE connection, try disconnecting and reconnecting |
| Pairing dialog appears but app won't proceed | You must tap **Pair** — canceling will block the connection |
---
## 11. LED Indicators (Device)
| LED State | Meaning |
|-----------|---------|
| Green blink (0.5s) | Bluetooth scanning |
| Green on 1s / off 3s | Device idle (not attached) |
| Orange blink (1s) | Sensor Alignment in progress |
| Green solid | Alignment complete (Optimal position) |
| Orange fast blink (3Hz) | Error detected |
| Orange solid | Charging |
| Blue solid | Charging complete |
---
## 12. Important Notes
- **Always use ultrasound gel** between the sensor and skin
- **Keep the sensor still** during measurement — movement reduces accuracy
- **Do not use the device while charging**
- The app keeps the screen on during use to prevent BLE disconnection
- Measurement logs are automatically saved to your device:
- BLE logs: `Downloads/VesiScan_BLE_*.log`
- ADC data: `Downloads/VesiScan_ADC/*.csv`
---
## Contact
For technical support, contact:
- App Development: dwjang
- Algorithm: Charles KWON / eunji.won
- Medithings Co., Ltd. — https://medithings.net